Inter-laboratory validation of procedures for measuring 8-oxo-7,8-dihydroguanine/8-oxo-7,8-dihydro-2 '-deoxyguanosine in DNA

2002 
The aim of ESCODD, a European Commission funded Concerted Action, is to improve the precision and accuracy of methods for measuring 8-oxo-7,8-dihydroguanine (8-oxoGua) or the nucleoside (8-oxodG). On two occasions, participating laboratories received samples of different concentrations of 8-oxodG for analysis. About half the results returned (for 8-oxodG) were within 20% of the median values. Coefficients of variation (for three identical samples) were commonly around 10%. A sample of calf thymus DNA was sent, dry, to all laboratories. Analysis of 8-oxoGua/8-oxodG in this sample was a test of hydrolysis methods. Almost half the reported results were within 20% of the median value, and half obtained a CV of less than 10%. In order to test sensitivity, as well as precision, DNA was treated with photosensitiser and light to introduce increasing amounts of 8-oxoGua and samples were sent to members. Median values calculated from all returned results were 45.6 (untreated), 53.9, 60.4 and 65.6 8-oxoGua/10 6 Gua;...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    3
    References
    69
    Citations
    NaN