Efficiency measurements performed on the MUSE VPHG
2010
Volume Phase Holographic Gratings (VPHG) are key elements for the second generation instrument MUSE (Multi Unit
Spectroscopic Explorer) developed for the VLT (Very Large Telescope) for ESO (European Southern Observatory).
MUSE operates in the visible wavelength range (465-930nm) and is composed of 24 spectrographs including one VPHG
each. This article briefly describes the design of the grating manufactured by Kaiser Optical Systems, to reach the MUSE
spectral resolution and efficiency. On the other hand the set up developed in CRAL (Centre de Recherche Astrophysique
de Lyon) to test the VPHG final performance is deeply discussed. This set up uses a broadband source coupled to a
monochromator, and a compensation arm to remove the source intensity fluctuations. The source is amplitude modulated
by a chopper, and a lock-in amplifier extracts the modulated signal from the photodiodes.
The measurement arm scans the 0, 1st and 2nd diffraction orders of the grating and allows tests of different areas over its
whole surface of 120mm*60mm. The accuracy reached is below one percent in efficiency, allows us to validate the
performance and its uniformity over the surface of the gratings.