Aptamer Functionalized Microcantilever Sensors for Cocaine Detection

2011 
A cocaine-specific aptamer was used as a receptor molecule in a microcantilever-based surface stress sensor for detection of cocaine molecules. An interferometric technique that relies on measuring differential displacement between two microcantilevers (a sensing/reference pair) was utilized to measure the cocaine/aptamer binding induced surface stress changes. Sensing experiments were performed for different concentrations of cocaine from 25 to 500 μM in order to determine the sensor response as a function of cocaine concentration. In the lower concentration range from 25 to 100 μM, surface stress values increased proportionally to coverage of aptamer/cocaine complexes from 11 to 26 mN/m. However, as the cocaine concentration was increased beyond 100 μM, the surface stress values demonstrated a weaker dependence on the affinity complex surface coverage. On the basis of a sensitivity of 3 mN/m for the surface stress measurement, the lowest detectable threshold for the cocaine concentration is estimated to...