Determination of Bisphenol A (BPA) in Plasma of Hemodialysis Patients Using Three Methods: LC/ECD, LC/MS, and ELISA

2008 
ABSTRACTBisphenol A (BPA) values were compared in plasma of hemodialysis patients and in recycling solvents using LC/MS, LC/electrochemical detector (ECD), and enzyme linked immunosorbent assay (ELISA). BPA values in the plasma and the solvent were 0–8.4 ng/ml and 0–0.8 ng/ml for LC/ECD, 0–4.9 ng/ml and 0–0.8 ng/ml for LC/MS, and 0–15.5 ng/ml and 0–3.1 ng/ml for ELISA, respectively. There was no significant difference among BPA values both in the plasma and the solvents using three methods. Single correlation coefficients between LC/ECD and LC/MS, LC/ECD and ELISA, and LC/MS and ELISA were, respectively, 0.373 (p 0.05), and 0.332 (p > 0.05) in the solvents (n = 19). An unknown substance present in the plasma of patients but not healthy volunteers influenced the LC/ECD values of plasma repeated freezing and thawing. The results indicate that LC/MS and ELISA are appropriate for BPA analysis in pla...