Calibration of imaging plate detectors to mono-energetic protons in the range 1-200 MeV

2017 
Responses of Fuji Imaging Plates (IPs) to proton have been measured in the range 1-200 MeV. Mono-energetic protons were produced with the 15 MV ALTO-Tandem accelerator of the Institute of Nuclear Physics (Orsay, France) and, at higher energies, with the 200-MeV isochronous cyclotron of the Institut Curie—Centre de Protontherapie d’Orsay (Orsay, France). The experimental setups are described and the measured photo-stimulated luminescence responses for MS, SR, and TR IPs are presented and compared to existing data. For the interpretation of the results, a sensitivity model based on the Monte Carlo GEANT4 code has been developed. It enables the calculation of the response functions in a large energy range, from 0.1 to 200 MeV. Finally, we show that our model reproduces accurately the response of more complex detectors, i.e., stack of high-Z filters and IPs, which could be of great interest for diagnostics of Petawatt laser accelerated particles.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    17
    Citations
    NaN
    KQI
    []