In Situ Monitoring of Pb2+ Leaching from the Galvanic Joint Surface in a Prepared Chlorinated Drinking Water

2018 
A novel method using a micro-ion-selective electrode (micro-ISE) technique was developed for in situ lead monitoring at the water–metal interface of a brass-leaded solder galvanic joint in a prepared chlorinated drinking water environment. The developed lead micro-ISE (100 μm tip diameter) showed excellent performance toward soluble lead (Pb2+) with sensitivity of 22.2 ± 0.5 mV decade–1 and limit of detection (LOD) of 1.22 × 10–6 M (0.25 mg L–1). The response time was less than 10 s with a working pH range of 2.0–7.0. Using the lead micro-ISE, lead concentration microprofiles were measured from the bulk to the metal surface (within 50 μm) over time. Combined with two-dimensional (2D) pH mapping, this work clearly demonstrated that Pb2+ ions build-up across the lead anode surface was substantial, nonuniform, and dependent on local surface pH. A large pH gradient (ΔpH = 6.0) developed across the brass and leaded-tin solder joint coupon. Local pH decreases were observed above the leaded solder to a pH as low...