A Proposed In Vitro Methodology for Assessing the Accuracy of Three-Dimensionally Printed Dental Models and the Impact of Storage on Dimensional Stability

2021 
The objective of this study was to propose a standardised methodology for assessing the accuracy of three-dimensional printed (3DP) full-arch dental models and the impact of storage using two printing technologies. A reference model (RM) comprising seven spheres was 3D-printed using digital light processing (MAX UV, MAX) and stereolithography (Form 2, F2) five times per printer. The diameter of the spheres (n = 35) represented the dimensional trueness (DT), while twenty-one vectors (n = 105) extending between the sphere centres represented the full-arch trueness (FT). Samples were measured at two (T1) and six (T2) weeks using a commercial profilometer to assess their dimensional stability. Significant (p 0.05) deviations were detected for FT. The detected median deviations were evenly distributed across the arch for MAX at <50 µm versus F2, where the greatest error of 278 µm was in the posterior region. Storage did not significantly impact the model’s DT in contrast to FT (p < 0.05). The proposed methodology was able to assess the accuracy of 3DP. Storage significantly impacted the full-arch accuracy of the models up to 6 weeks post-printing.