A rapid reversed phase high performance liquid chromatographic method for determination of etoposide (VP-16) in human plasma.

2001 
Abstract A rapid, simple and sensitive isocratic High Performance Liquid Chromatography (HPLC) method was developed to measure the concentration of etoposide in plasma samples with UV detection at 220 nm. The method uses a Bondapac C18 column at 60°C. The mobile phase consists of Methanol: water (45:55 v/v) at a flow rate of 2.8 ml/min. Phenacetin was used as an internal standard. The plasma samples were extracted using ether with the organic layer evaporated under nitrogen. The residue was dissolved in 200 μl methanol with 20 μl injected into the HPLC column. The extraction method showed a recovery of 91.5±3% for etoposide. In this system, the retention time of phenacetin and etoposide were 3.3 and 4.4 min, respectively. The limit of detection of etoposide in plasma is 20 ng/ml and the limit of quantitation is 40 ng/ml. This analytical method has very good reproducibility (8.1% between-day variability at a concentration of 50 ng/ml). It is a fast, sensitive and economic method applicable for clinical and pharmacokinetic studies.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    9
    References
    26
    Citations
    NaN
    KQI
    []