Advancement of X-ray radiography using microfocus X-ray source in conjunction with amplitude grating and SOI pixel detector, SOPHIAS
2018
We show how to improve microfocus X-ray radiography by using the SOPHIAS silicon-on-insulator pixel detector in conjunction with an amplitude grating. Single-exposure multi-energy absorption and differential phase contrast imaging was performed using the single amplitude grating method. The sensitivity in differential phase contrast imaging in a two-pixel-pitch setup was enhanced by 39% in comparison with the previously reported method [F. Krejci, Rev. Sci. Instrum.81, 113702 (2010).] by analyzing charge-sharing effects. Small-angle-scattering imaging was also possible in the two-pixel-pitch setup by counting the number of X-ray photons passing the pixel boundaries.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
3
Citations
NaN
KQI