Impact of screw diameter on pedicle screw fatigue strength - a biomechanical evaluation.

2021 
Objective Loosening of pedicle screws is a frequently observed complication in spinal surgery. Because additional stabilization procedures such as cement augmentation or lengthening of the instrumentation involve relevant risks, optimal stability of the primarily implanted pedicle screw is of essential importance. The aim of the present study was to investigate the effect of increasing the screw diameter on pedicle screw stability. Methods A total of 10 human cadaveric vertebral bodies (L4) were included in the present study. The bone mineral density was evaluated using quantitative computed tomography and the pedicle diameter using computed tomography. The vertebrae underwent instrumentation using 6.0-mm × 45-mm pedicle screws on 1 side and screws with the largest possible diameter (8–10-mm × 45-mm) on the other side. Fatigue testing was performed by applying a cyclic loading (craniocaudal sinusoidal 0.5 Hz) with increasing peak force (100 N + 0.1 N/cycle) until screw head displacement of 5.4 mm was reached. Results The mean fatigue load was 334 N for the 6-mm diameter screws and was increased significantly to 454 N (+36%) for the largest possible diameter screws (P  Conclusions Increasing the pedicle screw diameter from a standard 6-mm screw to the largest possible diameter (8–10 mm) led to a significantly greater fatigue load.