On-line characterization of gaseous and particulate organic analytes using atmospheric pressure chemical ionization mass spectrometry☆

2002 
Abstract A modified atmospheric pressure chemical ionization ion source is applied for direct analysis of volatile or low volatile organic compounds in air. The method is based on the direct introduction of the analytes in the gas phase and/or particle phase into the ion source of a commercial ion-trap mass spectrometer. Two methods are employed for the production of primary ions at atmospheric pressure, photoionization and corona discharge. It is shown that in the presence of a dopant, photoionization can be a highly efficient ionization method also for real-time analysis with detection limits for selected analytes in the lower ppt-range. Using corona discharge for the production of primary ions, which is instrumentally easier since no additional chemicals have to be added to the sample flow, we demonstrate the analytical potential of on-line atmospheric pressure chemical ionization mass spectrometry for reaction monitoring experiments. To do so, an atmospherically relevant gas phase reaction is carried out in a 500 l reaction chamber and gaseous and particulate compounds are monitored in the positive and negative ion mode of the mass spectrometer.