A realistic phantom for validating MRI‐based synthetic CT images of the human skull

2017 
Purpose To introduce a new realistic human skull phantom for the validation of synthetic CT images of cortical bone from ultra-short echo-time (UTE) sequences. Methods and Materials A human skull of an adult female was utilized as a realistic representation of skull cortical bone. The skull was stabilized in a special acrylic container and was filled with contrast agents that have T1 and T2 relaxation times similar to human brain. The phantom was MR scanned at 3T with UTE and T2-weighted sequences, followed by CT. A clustering approach was developed to extract the cortical bone signal from MR images. T2* maps of the skull were calculated. Synthetic CT images of the bone were compared to cortical bone signal extracted from CT images and confounding factors, such as registration errors, were analyzed. Results Dice similarity coefficient (DSC) of UTE-detected cortical bone was 0.84 and gradually decreased with decreasing number of spokes. DSC did not significantly depend on echo-time. Registration errors were found to be significant confounding factors, with 25% decrease in DSC for consistent 2mm error at each axis. Conclusion This work introduced a new realistic human skull phantom, specifically for the evaluation and analysis of synthetic CT images of cortical bone. This article is protected by copyright. All rights reserved.