Bio-analytical liquid chromatographic-based method with a mixed mode online solid phase extraction for drug monitoring of fluconazole in human serum

2021 
Abstract A simple, cost-effective and sensitive liquid chromatography-based bio-analytical method has been developed and validated for therapeutic drug monitoring of fluconazole (FLUC) in human serum. Integration of online mixed-mode solid-phase extraction (SPE) into the analytical system was the key for direct injection of untreated serum samples. A short protein-coated (PC) µBondapak CN silica column (PC-µB-CN-column) as a SPE tool and phosphate buffer saline (PBS) (pH 7.4) as an eluent were applied in the extraction step. PC-µB-CN-column operates in two different chromatographic modes. Using PBS, proteins were extracted from serum samples by size-exclusion liquid chromatography, while FLUC trapping was reversed-phase liquid chromatography dependent. FLUC was then eluted from the PC-µB-CN-column onto the quantification position using a mixture of acetonitrile-distilled deionized water (20:80, v/v) as an eluent and ODS analytical column. FLUC was separated at ambient temperature (22±1 °C) and detected at 260 nm. The method was linear over the range of 200-10000 ng/mL. FLUC recovery in untreated serum samples ranged from 97.8 to 98.8% and showed good accuracy and precision. The reliability of the developed method was evaluated by studying the pharmacokinetic profile of FLUC in humans after an oral administration of a single 150 mg tablet.