Optimization of chromatographic conditions for determination of aflatoxin B1, B2, G1 and G2 by using liquid chromatography-mass Spectrometry

2017 
The optimization of instrument condition and chromatographic separation for analysis of aflatoxin B1, B2, G1 and G2 using liquid chromatography tandem with mass spectrometer detector was conducted in the aim to provide more accurate and reliable analysis results. The aflatoxin known to be serious threat for human health as it is classified as the carcinogenic compounds. The aflatoxin B1, B2, G1 and G2 were selected due to its extensive contamination in various agricultural commodities. The best chromatographic separation was obtained using C-18 column with gradient elution of solvent 5 mM ammonium acetate and 0.1% formic acid in methanol at 7 minutes runtime analysis. The linearity of the detector showed satisfied results as the coefficient determination found to be 0.9994, 0.9996, 0.9998 and 0.9987 for aflatoxin B1, G1, B2, and G2 respectively in the range concentration from 1 to 20 ng/g. The quantifier ion selected for the aflatoxin B1, B2, G1 and G2 was m/z 285.1, 259, 243 and 313 respectively. The ins...