Analytical characterization of the APTIMA® HPV Assay

2009 
ABSTRACT Background Human papillomavirus (HPV) testing has improved the sensitivity for the detection of cervical pre-cancer and cancer as compared to Pap testing. Several HPV tests are commercially available and most target the DNA from 13 or 14 high-risk HPV types. The APTIMA® HPV Assay however, detects HPV E6/E7 mRNA from 14 high-risk types of HPV: 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68. Objective To determine the analytical performance characteristics of the APTIMA HPV Assay. Study Design Analytical sensitivity, analytical specificity, reproducibility, and the effect of potentially interfering substances was determined for the APTIMA HPV Assay on both the DTS (semi-automated) and TIGRIS DTS (fully automated) systems. Results The 95% detection limit for both systems was between 17 and 488 copies/reaction, depending on the HPV type. The assay did not cross-react with normal flora and opportunistic organisms that may be found in cervical samples, or low-risk HPV types. Spermicides, anti-fungal and anti-itch medications, whole blood, glacial acetic acid, and most lubricants did not interfere with assay performance. Those lubricants containing polyquaternium 15 did interfere with assay performance. Inter-instrument, inter-operator, inter-lot, and inter-run signal variability were 99% of the data. Intra-run variability was