A Tritiated-Water Detector with U-232/Th-228 Source
2001
The detection capabilities of the new U-232/Th-228 source are comparable to those of the Na-24 source. The main benefit in using the new source is the ease of operation. Elimination of the neutron activation step required for Na-24 sources saves about 24 hours in planning, scheduling, and executing. With the new U-232/Th-228 source, the monitor can be put in operation in less than 15 minutes. The long half-life of the U-232/Th-228 source also eliminates the need to record calibration and measurement times, as required for decay corrections when using a Na-24 source.