Performance of the Abbott RealTime HIV-1 Viral Load Assay Is Not Impacted by Integrase Inhibitor Resistance-Associated Mutations
2011
The Abbott RealTime HIV-1 viral load assay uses primers and probes targeted to integrase, which is also the target of integrase inhibitors such as raltegravir. Viral loads of 42 raltegravir-susceptible and 40 raltegravir-resistant specimens were determined using RealTime HIV-1 and Roche Monitor (v1.5). The differences in viral load measurements between assays were comparable in the two groups, demonstrating that the RealTime HIV-1 assay can tolerate raltegravir-selected mutations.