Optimization of direct analysis in real time (DART) linear ion trap parameters for the detection and quantitation of glucose

2012 
Presented here are findings for the development and optimization of a simple, high-throughput, and rapid method for the analysis of glucose. Because the applications of glucose and other six-carbon sugars is a growing field of interest especially in the production of biofuels, an efficient and rapid method for their quantitation from lignocelluloses is necessary. Glucose was analyzed using direct analysis in real time (DART) ionization and formed adducts (along with fragmentation) were observed with a linear ion trap (LIT) mass spectrometer. Since DART can be considered a complex thermal desorption ionization process, an optimization study of the helium gas temperature and introduction into the ionization region was performed. It was observed these parameters have a significant effect on the overall signal intensity as well as the signal-to-noise ratios in DART mass spectra. Using these optimized parameters, a set of different glucose concentrations (ranging from 10 to 3000 μM) were analyzed and used to determine a linear dynamic range (with the use of an internal standard). The analysis of the samples was done with minimal sample preparation and found to be reproducible on different days. Copyright © 2012 John Wiley & Sons, Ltd.