THE PERFORMANCE OF MIXED MANUFACTURER METAL ON METAL TOTAL HIP REPLACEMENTS

2018 
Using a femoral head from one manufacturer on the stem of another manufacturer poses the risk that the taper interface between the components do not contact correctly and the performance of the joint will be impaired. The cohorts in this study are a combination of modular Birmingham Hip Resurfacing (BHR) and Adept femoral heads on CPT stems. The study reviews the geometry of the taper interfaces to establish if the taper clearance angles was outside of the normal range for other taper interfaces. In addition the rates of material loss from the bearings and taper and a ranking of the stem damage were reviewed to determine if the levels of loss were above that seen for other similar joints.The material loss analysis demonstrated that the rates or levels of loss from the bearings, taper and stem are no different to levels published for manufacturer matched joints and in many cases are lower. The results demonstrate that the taper clearance angles for the mixed manufacturer joints (BHR-CPT: 0.067 to −0.116, A...