The calibration and intercalibration of sea-going infrared radiometer systems using a low cost blackbody cavity

1999 
Abstract There are many infrared radiometer systems available for the measurement of in situ sea surface skin temperature (SSST). Unfortunately, the marine environment is extremely hostile to optical components, and to ensure the accuracy of SSST measurements, an absolute calibration of instrumentation using an independent calibration reference is required both before and after any sea deployment. During extended deployments it is prudent to have additional regular calibration data to monitor instrument performance characteristics. This paper presents a design for an ambient temperature (278–325 K), wide aperture (100 mm), reference blackbody unit that may be used to calibrate a variety of sea-going infrared radiometer systems both in the laboratory and in the field. The blackbody consists of a spun copper cavity coated with well-characterized high emissivity paint (Mankiewicz Nextel Velvet Coating 811-21) immersed in a water bath that is continuously mixed using a strong water pump. The radiant temperatu...