Counting and Integrating Microelectronics Development for Direct Conversion X-ray Imaging

2008 
A novel signal processing concept for X-ray imaging with directly converting pixelated semiconductor sensors is presented. The novelty of this approach compared to existing concepts is the combination of charge integration and photon counting in every single pixel. Simultaneous operation of both signal processing chains extends the dynamic range beyond the limits of the individual schemes and allows determination of the mean photon energy. Medical applications such as X-ray computed tomography can benefit from this additional spectral information through improved contrast and the ability to determine the hardening of the tube spectrum due to attenuation by the scanned object. A prototype chip in 0.35-micrometer technology has been successfully tested. The pixel electronics are designed using a low-swing differential current mode logic. Key element is a configurable feedback circuit for the charge sensitive amplifier which provides continuous reset, leakage current compensation and replicates the input signal for the integrator. The electronic characterization of a second generation prototype chip is described and a detailed discussion of the circuit design is given.