Femtosecond broadband fluorescence upconversion spectroscopy: Spectral coverage versus efficiency

2016 
Sum frequency mixing of fluorescence and ∼1300 nm gate pulses, in a thin β-barium borate crystal and non-collinear type II geometry, is quantified as part of a femtosecond fluorimeter [X.-X. Zhang et al., Rev. Sci. Instrum. 82, 063108 (2011)]. For a series of fixed phasematching angles, the upconversion efficiency is measured depending on fluorescence wavelength. Two useful orientations of the crystal are related by rotation around the surface normal. Orientation A has higher efficiency (factor ∼3) compared to B at the cost of some loss of spectral coverage for a given crystal angle. It should be used when subtle changes of an otherwise stationary emission band are to be monitored. With orientation B, the fluorescence range λF > 420-750 nm is covered with a single setting of the crystal and less gate scatter around time zero. The accuracy of determining an instantaneous emission band shape is demonstrated by comparing results from two laboratories.