Comparison of biologically effective dose for treatment planning in the fixed-beam intensity-modulated radiotherapy and the volumetric-modulated arc therapy for the typical types of cancer

2019 
Abstract Purpose To compare the fixed-beam intensity-modulated radiotherapy (IMRT) with the volumetric-modulated arc therapy (VMAT) plans for different cancers using the biologically effective dose (BED). Methods Recruited patients were divided into three groups in terms of tumor anatomic site, including head and neck, thoracic, and abdominal cancers. All of plans were generated using the fixed-beam IMRT and the VMAT techniques. The qualities of the plans were assessed using the conformity index (CI) and the homogeneous index (HI) with respect to the dose delivered to the tumors. Some selected organs-at-risk (OARs) were analyzed. For the efficiency of the plan, monitor units (MUs) were compared between the paired plans. All parameters were collected and evaluated based on the conventional dose-volume histogram (DVH) of the paired plans. All BED were calculated from the physical dose. Results For CI and HI, the fixed-beam IMRT and the VMAT gained significant differences ( p p p Conclusion There is no difference in the data comparison of the paired plans using the BED values compared with the physical doses.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    30
    References
    0
    Citations
    NaN
    KQI
    []