Verifying the validity of creatinine measurement in low‐concentrated urine spot samples ‐ photospectrometry vs. liquid chromatography‐tandem mass spectrometry

2021 
One of the major challenges of testing drugs of abuse is the detection of highly diluted urine samples. The ingestion of a large amount of fluid can considerably reduce the concentration of substances, possibly resulting in inaccurate drug testing. For detection, determination of urinary creatinine is a widely established procedure. In this study, results from the most popular methods, including photospectrometry (Jaffe) and liquid chromatography-tandem mass spectrometry (LC-MS/MS), have been compared regarding 327 urine abstinence control samples. Because samples with creatinine concentrations close to the cutoff of 20 mg/dL are of particular interest, only samples below 50 mg/dL were considered. Results revealed a close correlation of creatinine concentrations by both analytical methods with an R2 value of 0.9005. A mean concentration difference of 3.30 ± 3.45 mg/dL was observed, indicating a moderate underestimation by the Jaffe reaction. Graphical analyses showed high accordance between both methods with only a few outliers. Due to easy handling and for economic reasons, the spectrometric method is often preferred over LC-MS/MS. For urine samples with creatinine concentrations close to the cutoff, confirmation through a second method should be performed to avoid a possible disadvantage or even severe consequences for the respective individual. It is recommended that each laboratory establishes a reliable verification method.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    1
    Citations
    NaN
    KQI
    []