Systematic evaluation of magnetic sensitivities of anisotropic magnetoresistive sensors at liquid helium temperature for superconducting cavities

2021 
Trapped magnetic flux in bulk superconductors reduces the quality factor Q in superconducting radio-frequency (SRF) cavities. However, the mechanisms underlying flux trapping and radio-frequency loss are not well understood. Detailed observation of the magnetic distributions is important for understanding such phenomena. Magnetic field mapping is useful for observing the magnetic field distribution around SRF cavities. Measuring the change in the magnetic field around the cavity elucidates the flux trapping behavior. Anisotropic magnetoresistive (AMR) sensors are inexpensive and small devices that can detect magnetic flux density. The magnetic sensitivities of AMR sensors need to be evaluated at liquid helium temperature for the magnetic field mapping of SRF cavities. In this study, a test stand was constructed to calibrate the magnetic sensitivities of AMR sensors in liquid helium, and 110 AMR sensors were tested using this stand. The magnetic sensitivities were evaluated systematically. A solenoid coil was used to control the uniform external magnetic field and to measure the magnetic sensitivity at low temperatures. All AMR sensors exhibited suitable sensitivities to the magnetic field around the SRF cavity. The variation in these sensitivities in all AMR sensors was ∼1%. The AMR sensors were found to have sufficient sensitivity for mapping the magnetic field around the exterior surface of the SRF cavity.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17