THE NEW CHALK RIVER AMS ION SOURCE, SAMPLE CHANGER AND EXTERNAL SAMPLE MAGAZINE

1997 
Abstract A new sample magazine, sample changer and ion source have been developed and are in routine use at Chalk River. The system features a readily accessible 40-sample magazine at ground potential that is external to the ion source and high-voltage cage. The samples are held in an inert atmosphere and can be individually examined or removed; they can be exchanged en masse as a complete magazine concurrent with an AMS measurement. On-line sample changing is done with a pneumatic rabbit transfer system employing two stages of differential pumping. At Chalk River this is routinely performed across a 200 kV potential. Sample positioning is precise, and hundreds of 36 Cl and 129 I samples have been measured over a period of several days without interruption or alteration of ion source operating conditions.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    6
    Citations
    NaN
    KQI
    []