Present status of 920 MHz high-resolution NMR spectrometers

2004 
The first 920 MHz high-resolution NMR (nuclear magnetic resonance) magnet was successfully installed at the Tsukuba Magnet Laboratory of the National Institute for Materials Science. A persistent operation at 21.6 T has continued since April 2002 without any problems. Excellent field stability of 0.31 Hz/h was observed from December 2002 to January 2003. Since July 2002, the magnet was used as an essential part of the only 920 MHz high-resolution NMR spectrometer in the world. The signal-to-noise ratio of 0.1% ethylbenzene using a proton-selective probe was as high as 2981. An HCN triple-resonance probe is now attached to the spectrometer to contribute to the National Project on Protein Structural and Functional Analysis in Japan. The completion of the second 920 MHz-class NMR spectrometer is scheduled for spring 2004.