Design Considerations and Performance Evaluation of a Compact Aerosol Particle Mass Analyzer

2013 
A compact aerosol particle mass analyzer (APM) of which the size of the classifier was significantly reduced than that of the first commercial model (Kanomax Model 3600) was developed. Firstly, requirements for desired performance in classifying particle mass were set forth. Secondly, a theoretical framework for the design parameters of an APM that satisfies the requirements was formulated. Thirdly, the design parameters were determined that satisfies the requirements while reducing the instrument size. The requirements include the condition that the classification range covers from 0.001 to 1000 fg (approximately 12 to 1200 nm in size for spherical particles having the density of 1 g/cm3), and the condition that both the classification resolution and particle penetration in this mass range are higher than certain specified values. A prototype having the design parameters determined according to this theoretical framework was constructed, and its performance was evaluated experimentally. The external dime...