Detection of moderate hyperhomocysteinemia: Comparison of the Abbott fluorescence polarization immunoassay with the Bio-Rad and SBD-F high-performance liquid chromatographic assays Short Communication
2001
Summary. The importance of accurate methods for homocysteine measurement has been emphasized. We compared the results obtained with the most commonly used high-performance liquid chromatography (HPLC) assay, and two recently commercially available methods: another HPLC and a fluorescence polarization immunoassay, in plasmas from normo- or hyperhomocysteinemic patients. A significant agreement between the different methods in classifying the results as hyper or normal-homocysteinemia was observed. However, a significant difference between the results was found. Standardization is urgently necessary to improve the concordance of homocysteine determination.