Predictive gamma passing rate by dose uncertainty potential accumulation model

2018 
PURPOSE: Intensity-modulated radiation therapy (IMRT) utilizes many small fields for producing a uniform dose distribution. Therefore, there are many field junctions in the target region, and resulting dose uncertainties are accumulated. However, such accumulation of the dose uncertainty has not been implemented in the current practice of IMRT dose verification. The purpose of this study is to develop a method to predict the gamma passing rate (GPR) using a dose uncertainty accumulation model. METHODS: Thirty-three intensity-modulated (IM) beams for head-and-neck cases with step-and-shoot techniques were used in this study. The treatment plan was created using the XiO treatment planning system (TPS). The IM beam was produced by the ONCOR Impression Plus linear accelerator. MapCHECK was used to measure the dose distribution. The distribution of a dose uncertainty potential (DUP) was generated by in-house software that accumulated field shapes weighted by a segmental monitor unit, followed by Gaussian folding. The width of the Gaussian was determined from the width of the lateral penumbra. The dose difference between the calculated and measured doses was compared with the estimated DUP at each point. The GPR of each beam was predicted for 2%/2-mm, 3%/2-mm, and 3%/3-mm tolerances by its own DUP histogram and a GPR-vs-DUP correlation of other beams using the leave-one-out cross-validation method. The predicted GPR was compared with the measured GPR to evaluate the performance of this prediction method. The criteria for the predicted GPR corresponding to a measured GPR ≥ 90% were estimated to examine the feasibility of estimating the measured GPR by this GPR prediction method. RESULTS: The DUP was confirmed to have proportionality to the standard deviation (SD) of the dose difference. The SDs of the difference between the measured and predicted GPRs were 3.1, 1.7, and 1.4% for 2%/2-mm, 3%/2-mm, and 3%/3-mm tolerances, respectively. The criteria of the predicted GPR corresponding to the measured GPR ≥ 90% were 94.1 and 95.0% with confidence levels of 99 and 99.9%, respectively. CONCLUSION: In this study, we confirmed the good proportionality between the dose difference and the estimated DUP. The results showed a feasibility to predict the dose difference from DUP as estimated by a DUP accumulation model. The predicted GPR developed in this study showed good accuracy for planar dose distributions of head and neck IMRT. The prediction method developed in this study is considered to be feasible as a substitute for the current practice of measurement-based verification of the dose distribution with gamma analysis.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    5
    Citations
    NaN
    KQI
    []