Design of Cone Magnets and Shielding to Align and Calibrate Hall Probe Measurement System

2018 
To measure the magnetic fields with a Hall probe, accurate alignment needs to be considered. Therefore, we present improvements to the Hall probe measurement system, including cone magnets and a zero-field shield as simulated by the OPERA. The design of the cone magnets of the North-face-North type is greatly helpful to monitor the magnet alignment. The accurate location of the Hall probe doping area should be determined before doing any measurements. In addition, a zero-field shielding tube is used to reset the probe to zero. The results of simulations are reported in this paper.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    7
    References
    0
    Citations
    NaN
    KQI
    []