Investigating critical flicker fusion frequency for monitoring gas narcosis in divers

2020 
Introduction Critical flicker fusion frequency (CFFF) has been used in various studies to measure the cognitive effects of gas mixtures at depth, sometimes with conflicting or apparently paradoxical results. This study aimed to evaluate a novel automatic CFFF method and investigate whether CFFF can be used to monitor gas-induced narcosis in divers. Methods Three hyperbaric chamber experiments were performed: 1) Automated and manual CFFF measurements during air breathing at 608 kPa (n = 16 subjects); 2) Manual CFFF measurements during air and heliox breathing at sea level (101.3 kPa) and 608 kPa (n = 12); 3) Manual CFFF measurements during oxygen breathing at sea level, 142 and 284 kPa (n = 10). All results were compared to breathing air at sea level. Results Only breathing oxygen at sea level, and at 284 kPa, caused a significant decrease in CFFF (2.5% and 2.6% respectively compared to breathing air at sea level. None of the other conditions showed a difference with sea level air breathing. Conclusions CFFF did not significantly change in our experiments when breathing air at 608 kPa compared to air breathing at sea level pressure using both devices. Based on our results CFFF does not seem to be a sensitive tool for measuring gas narcosis in divers in our laboratory setting.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    2
    Citations
    NaN
    KQI
    []