Development and validation of a method for gefitinib quantification in dried blood spots using liquid chromatography-tandem mass spectrometry: Application to finger-prick clinical blood samples of patients with non-small cell lung cancer

2018 
Abstract A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated for the quantification of gefitinib in dried blood spots (DBSs). Gefitinib was extracted with methanol from DBS of 3 mm in diameter and detected using a triple quadrupole mass spectrometer. The method was validated by evaluating its precision, accuracy, selectivity, carryover, matrix effect, recovery, and stability. For clinical validation, paired finger-prick DBS and plasma concentrations were compared for 10 patients with non-small cell lung cancer (NSCLC) taking gefitinib. The calibration linear range was 37.5–2400 ng/mL (coefficient of determination [R 2 ] = 0.99), encompassing the therapeutic concentrations of gefitinib. The accuracy and precision were within 15% of the quality control (QC) concentrations of 80, 200, and 2000 ng/mL. The lower limit of quantification was determined to be 40 ng/mL. Gefitinib was stable in DBSs for up to 5 months at room temperature and −20 °C, and at 40 °C for 24 h. A good correlation was observed between the gefitinib levels measured by the DBS method and plasma concentrations (R 2  = 0.99). This method provides a simple, fast, and accurate approach to the quantitative analysis of gefitinib in finger-prick DBSs. The method would be useful for minimally invasive evaluation of the clinical gefitinib blood concentration.