Feasibility and Performance of Loop-Mediated Isothermal Amplification Assay in the Diagnosis of Pulmonary Tuberculosis in Decentralized Settings in Eastern China

2019 
Early diagnosis is essential for the control and prevention of tuberculosis (TB). The objective of this study was to investigate the feasibility and performance of loop-mediated isothermal amplification (LAMP) in the diagnosis of pulmonary TB in county-level microscopy centers in Qingdao, Eastern China. A total of 523 presumptive TB patients were consecutively recruited between July 2017 and April 2018, and 22 patients were excluded from the analysis. Of 102 culture-positive cases, TB-LAMP identified 91 cases, demonstrating a sensitivity of 89.2%. In comparison, the sensitivity of routine smear microscopy was 69.6% (71/102), which was significantly lower than that of TB-LAMP (P=0.001). In addition, TB-LAMP sensitivities in smear-positive and smear-negative samples were 98.6% and 67.7%, respectively. In conclusion, our data demonstrate that TB-LAMP outperforms conventional smear microscopy in TB diagnosis, which could be used as an alternative method for smear microscopy in resource-limited settings in China.