Step-length biofeedback device for walk rehabilitation

1994 
A biofeedback gait training system for step length is proposed, adapted to the correction of spatial walking asymmetries by means of a simple, quick and reliable method for daily clinical use. The system is composed of a walkway and a gait analysis device (locometer) measuring the main temporal and distance factors of gait. The step length is imposed on the subject by lighted targets appearing on the walkway, alternately on the right and left side; the subject is asked to place a swinging foot on the lighted target. Feedback to the subject is supplied by direct visual information (the subject looking at the movement and the position of the foot with respect to the lighted target) and an acoustic signal delivered in real time when the length step error is greater than an allowed value. The method is validated on a population of hemiparetic patients who have suffered from a stroke and who have been reeducated with traditional rehabilitation methods. The patients were divided into two groups; one group following a gait training with biofeedback (BFB group)_ and one group following a gait training without biofeedback (reference group). Preliminary results are presented, showing a significant beneficial effect of the biofeedback method in increasing the step length of paretic limbs and in correcting step-length asymmetry.