Optimization of an enzyme-coupling method by spectrophotometer for serum adenosine deaminase: As a candidate reference method
2019
Abstract Adenosine deaminase (ADA) is a key enzyme of adenosine metabolism. There are currently various kits and systems available for ADA measurement, and all yield variable results. This study optimized a reference measurement procedure (RMP) for serum ADA for the standardization of routine methods. ADA coupled with purine-nucleoside-phosphorylase, xanthine-oxidase and peroxidase was selected as the basic method and was optimized using Response Surface Methodology. Then the performance was validated and the results were compared after replication by 3 other reference laboratories. A reference interval was also developed. In addition, this optimized method was applied to calibrate a routine system. The intra-assay precision was 0.44% at both concentrations of 29.8 and 100.4 U/L, and inter-assay precision was 1.01% and 0.95% at 30.1 and 100.3 U/L, respectively. The linearity was up to 351.9 U/L (R2 = 0.9998), with no significant interference or carryover (
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
22
References
0
Citations
NaN
KQI