Noise adding radiometer performance analysis

1980 
The DSN noise adding radiometer (NAR) measurement accuracy is analyzed. The NAR capability is part of the precision power monitor function recently introduced in the DSN. The potential system noise temperature measurement accuracy is estimated to be about 1.5% with the 1 kelvin noise diode. Performance verification requires comparison noise temperature measurements with the manual Y factor method and routine monitoring of critical elements in the NAR system such as system linearity and noise diode calibration. A technique is presented to calibrate and reduce the effects of the receiving system nonlinearity. Unsatisfactory performance or degradation of these critical NAR elements would require appropriate system upgrading.