Digital holography microscopy (DHM) operating in pulsed stroboscopic mode: a versatile metrology instrument for micro and nano technology
2007
Process engineering and failure analysis of MEMS and MOEMS require static and dynamical characterization of both
their in-plane and out of plane response to an excitation. A remarkable characteristic of Digital Holography Microscopes
(DHM) is the extremely short acquisition time required to grab the whole information necessary to provide 3D optical
topography of the sample: a unique frame grab, without any vertical or lateral scan provides the information over the
full field of view. First, it ensures DHM measurements to be insensitive to vibrations. Second, it opens the door to fast
dynamical characterization of micro-systems. For periodic movement analysis, DHM can operate in two stroboscopic
modes with standard cameras. The first one enables precise characterization up to excitation frequencies of 100 kHz
with recovery cycle of 10% simply by triggering properly the camera. The second one uses a pulsed source for
investigation of higher excitation frequencies. For non periodic movement analysis fast acquisition cameras and
postponed treatment are used. DHM are therefore unique and very efficient tool for dynamical characterization of inhplane
and out-of-plane response.
In this paper we illustrate the two stroboscopic modes with an example of a high frequency micro mirror.