Hydrodynamic Flows Around Tablets in Different Pharmacopeial Dissolution Tests

2002 
ABSTRACTWe investigated the hydrodynamic flows around tablets during several pharmacopeial dissolution tests: the rotating basket (RB), paddle (PD), flow-through cell (FT), and disintegration (DI) tests. The determination of hydrodynamic flow was based on the dissolution rate of United States Pharmacopeial salicylic acid nondisintegrating calibrators, and showed that, compared with the PD and RB methods, the FT method produced a lower hydrodynamic flow value whereas the DI method produced a higher value. The hydrodynamic flows during the PD and RB tests appeared to be similar at the same rotational speed, although the flow patterns around the tablet differed; with the RB method, homogeneous dissolution occurred from all surfaces of the tablet, while with the PD method, dissolution from the lower surface was slower. The use of a sinker seemed to enhance dissolution from the lower surface. Such differences in hydrodynamic flow could explain the apparently different dissolution behaviors of disintegrating pr...