Minimizing scattering-induced phase errors in differential interference contrast microscopy

2020 
Significance Differential interference contrast (DIC) microscopes allow noninvasive in vivo observation of transparent microstructures in tissue without the use of fluorescent dyes or genetic modification. We show how to modify a DIC microscope to measure the sample phase distribution accurately and in real-time even deep inside sample tissue. Aim Our aim is to improve the DIC microscope's phase measurement to remove the phase bias that occurs in the presence of strong scattering. Approach A quarter-wave plate was added in front of the polarization camera, allowing a modified phase calculation to incorporate all four polarization orientation angles (0 deg, 45 deg, 90 deg, and 135 deg) captured simultaneously by the polarization camera, followed by deconvolution. Results We confirm that the proposed method reduces phase measurement error in the presence of scattering and demonstrate the method using in vivo imaging of a beating heart inside a medaka egg and the whole-body blood circulation in a young medaka fish. Conclusions Modifying a polarization-camera DIC microscope with a quarter-wave plate allows users to image deep inside samples without phase bias due to scattering effects.