Characteristics of a Frisch collar grid CdZnTe radiation detector grown by low-pressure Bridgman method

2015 
A single-polarity charge-sensing method was studied by using a CdZnTe Frisch collar grid detector grown by using a low-pressure Bridgeman furnace at the Korea Atomic Energy Research Institute (KAERI). The Frisch collar grid CdZnTe detector has an active volume of 5 × 5 × 10 mm3, and was fabricated from a single crystal, Teflon tape and copper tape used as a Frisch collar grid. A room-temperature energy resolution of 6% full width at half maximum (FWHM) was obtained for the 662keV peak of Cs-137 without any additional electrical corrections. The detector’s fabrication process is described, and its characteristics are discussed. Finally, the charge transport properties and gamma-ray energy resolution of the fabricated Frisch collar grid detector are compared with those of two other Frisch collar detectors grown by using different crystal growth methods and purchased from eV-products and Redlen technology.