The gel test is the most prevalent method for the forward and reverse blood typing tests. It relies on the controlled centrifugation of red blood cells (RBCs) and antibodies through a gel column. This noncontinuous matrix is currently based on microbeads that often lack sensitivity. For the first time, nanocellulose hydrogel is demonstrated as a sustainable and reliable medium for gel-based blood typing diagnostics. Gels with a minimum of 0.3 wt % TEMPO-oxidized cellulose nanofibers (0.92 mmol/g of carboxyl content) separate agglutinated and individual RBCs in the forward test. The addition of glycine is able to balance the osmotic pressure and reduce hemolysis to 5%, while retaining the electrostatic repulsion responsible for the gel network structure and its rheological properties. For the reverse typing, cellulose nanofibers are chemically cross-linked with hexamethylenediamine (HMDA), increasing the gel yield point 8-fold. Sodium chloride is added to achieve the osmolality found in the human plasma an...