InstantLabs® listeria species food safety kit

2014 
: The InstantLabs Listeria Species Food SafetyKitwas validated against the International Organization for Standardization (ISO) reference method 11290-1 for the detection of Listeria monocytogenes and other Listeria species. The matrixes (stainless steel, sealed concrete, cheddar cheese, raw shrimp, and hot dogs) were inoculated with approximately 1 CFU/test portion of various Listeria species to generate fractional positives (5-15) in 20 inoculated samples. Enrichments were also fractionally inoculated with L. monocytogenes for side-by-side testing of the InstantLabs Listeria monocytogenes Food Safety Kit. Stainless steel and sealed concrete samples were validated using 4 x 4" and 1 x 1" test areas, respectively, and enriched in Buffered Listeria Enrichment Broth (BLEB) at 35 +/- 1 degrees C for 22-28 h. All food samples were tested at 25 g or 25 mL and enriched in BLEB at 35 +/- 1 degrees C for 24-28 h. All samples were confirmed using the ISO reference method, regardless of initial screen result. The InstantLabs test method performed as well as or better than the reference method for the detection of Listeria species on stainless steel, sealed concrete, cheddar cheese, raw shrimp, and hot dogs. Inclusivity and exclusivity testing revealed no false negatives and no false positives among the 80 Listeria species and 30 non-Listeria species examined. The method was shown to be robust when variations were introduced to the enrichment time, the volume for DNA extraction, and the heat block time (data not shown).