The accuracy of personal activity monitoring devices

2017 
Abstract Activity monitoring has important applications in orthopaedic and medical research. There is a paucity of information on device accuracy. Thirty adults tested the accuracy of 10 devices in 5 activities: walk 400-M, run 400-M, walk 10-M, ascend and descend 10 steps. A second protocol tested slow walking speeds at 1.0, 1.5, 1.8, and 2.0MPH. The waist-based devices FitBit One and Omron HJ-321 were >90% accurate for all activities. The wristband devices and Smartphone Apps were 95% accurate at lower cadence activities, but undercounted running by ~25%. Waist-based, dedicated activity monitors are highly accurate in a variety of activities.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    14
    Citations
    NaN
    KQI
    []