Reflex Testing Rules for Cell Count and Differentiation of Nucleated Elements in Pleural and Ascitic Fluids on Sysmex XE-5000:

2016 
Flow cytometry is widely used in many laboratories for automated nucleated cell counts and their differentiation in body fluids. The implementation of new reflex testing rules on these automated instruments could open new frontiers in laboratory workflow, improving characterization of body fluids and clinical diagnosis and decreasing costs. Ascitic (150) and pleural (33) fluids were collected and assessed by XE-5000 and optical microscopy. Cell counts performed with the methods showed a Pearson’s correlation of 0.98 (p < 0.0001), Passing–Bablok regression y = 0.99x + 2.44, and bias of 32.3. In ascitic fluids, the best diagnostic performance was found for polymorphonuclear and neutrophil counts on XE-5000, which exhibited areas under the curve (AUCs) 0.98 (p < 0.0001) and 0.99 (p < 0.0001), respectively. In pleural fluids the best diagnostic performance was found for polymorphonuclear percent parameter, which displayed 0.97 (p < 0.0001). Specific reflex test rules based on these parameters were characteriz...