Cell micro-irradiation with MeV protons counted by an ultra-thin diamond membrane

2017 
We report the development of thin single crystal diamond membranes suitable for dose control in targeted cell irradiation experiments with a proton microbeam. A specific design was achieved to deliver single protons with a hit detection efficiency approaching 100%. The membranes have thicknesses between 1.8 and 3 μm and are used as vacuum windows on the microbeam line. The impact of these transmission detectors on the microbeam spot size is estimated by Monte-Carlo simulations, indicating that a beam lateral resolution below 2 μm is achieved. This is confirmed by experiments showing the accumulation online of X-ray Repair Cross-Complementing protein 1 (XRCC1)-Green Fluorescent Protein (GFP) at DNA damaged sites in living cells.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    3
    Citations
    NaN
    KQI
    []