A new method for output factor determination in MLC shaped narrow beams

2007 
Abstract A new method for the measurement of output factors of narrow beams is presented in this work. By combining a new large area parallel plane ionization chamber (PTW model T34070) with a relative film dosimetry the output factors of small square fields of a 6 MV beam shaped by a MLC were measured. Several detectors (three ionization chambers, two solid state detectors and film) and Monte Carlo simulation were also employed to validate this new methodology and also to determine those detectors more suitable for narrow beam output factor determination. The proposed method for output factor measurement has shown to be in a very good agreement with diamond, diode and Monte Carlo results while it is insensitive to position displacements. Several uncertainties associated to the process of narrow beam dosimetry have also been addressed.