Quantifying Soft Tissue Artefacts and Imaging Variability in Motion Capture of the Fingers

2020 
This study assessed the accuracy of marker-based kinematic analysis of the fingers, considering soft tissue artefacts (STA) and marker imaging uncertainty. We collected CT images of the hand from healthy volunteers with fingers in full extension, mid- and full-flexion, including motion capture markers. Bones and markers were segmented and meshed. The bone meshes for each volunteer’s scans were aligned using the proximal phalanx to study the proximal interphalangeal joint (PIP), and using the middle phalanx to study the distal interphalangeal joint (DIP). The angle changes between positions were extracted. The HAWK protocol was used to calculate PIP and DIP joint flexion angles in each position based on the marker centroids. Finally the marker locations were ‘corrected’ relative to the underlying bones, and the flexion angles recalculated. Static and dynamic marker imaging uncertainty was evaluated using a wand. A strong positive correlation was observed between marker- and CT-based joint angle changes with 0.980 and 0.892 regression slopes for PIP and DIP, respectively, and Root Mean Squared Errors below 4°. Notably for the PIP joint, correlation was worsened by STA correction. The 95% imaging uncertainty interval was < ± 1° for joints, and < ± 0.25 mm for segment lengths. In summary, the HAWK marker set’s accuracy was characterised for finger joint flexion angle changes in a small group of healthy individuals and static poses, and was found to benefit from skin movements during flexion.