Aim:  To determine the performance of the Ridascreen® SET Total kit, after sample extraction and concentration by dialysis, with regard to its use in official controls for staphylococcal enterotoxins under European Regulation (EC) No. 2073/2005 modified. This study was conducted on naturally contaminated cheese samples and compared with the results of the previously validated Vidas® SET2 kit. Methods and Results:  The effectiveness of the Ridascreen® SET Total kit on naturally contaminated cheeses was compared to that of the Vidas® SET2 kit by applying the EN ISO 16140 standard. Sensitivity and specificity were also compared using spiked buffer solutions and cheese samples with SEA to SEE toxins. Conclusions:  This study showed that the Ridascreen® SET Total kit is as effective as the Vidas® SET2 kit. Significance and Impact of the Study:  The Ridascreen® SET Total kit was found to specifically detect SEA to SEE in cheeses. The Ridascreen® SET Total can therefore be used to check the staphylococcal enterotoxin content and ensure consumer protection.