Development and performance validation of a cryogenic linear stage for SPICA-SAFARI verification
2014
In the context of the SAFARI instrument (SpicA FAR-infrared Instrument) SRON is developing a test environment to
verify the SAFARI performance. The characterization of the detector focal plane will be performed with a backilluminated
pinhole over a reimaged SAFARI focal plane by an XYZ scanning mechanism that consists of three linear
stages stacked together. In order to reduce background radiation that can couple into the high sensitivity cryogenic
detectors (goal NEP of 2•10 -19 W/√Hz and saturation power of few femtoWatts) the scanner is mounted inside the
cryostat in the 4K environment. The required readout accuracy is 3 μm and reproducibility of 1 μm along the total travel
of 32 mm. The stage will be operated in “on the fly” mode to prevent vibrations of the scanner mechanism and will
move with a constant speed varying from 60 μm/s to 400 μm/s. In order to meet the requirements of large stroke, low
dissipation (low friction) and high accuracy a DC motor plus spindle stage solution has been chosen.
In this paper we will present the stage design and stage characterization, describing also the measurements setup. The
room temperature performance has been measured with a 3D measuring machine cross calibrated with a laser
interferometer and a 2-axis tilt sensor. The low temperature verification has been performed in a wet 4K cryostat using a
laser interferometer for measuring the linear displacements and a theodolite for measuring the angular displacements.
The angular displacements can be calibrated with a precision of 4 arcsec and the position could be determined with high
accuracy. The presence of friction caused higher values of torque than predicted and consequently higher dissipation.
The thermal model of the stage has also been verified at 4K.